Technical Specifications; Technical Specifications Ultramat 23 - Siemens ULTRAMAT 6 Series Compact Operating Instructions

Continuous gas analysis
Table of Contents

Advertisement

Technical specifications

7.1

Technical specifications ULTRAMAT 23

General information
Measured components
Measuring ranges
Characteristics
Operator panel
Operating position
Vibrations
During operation
During transport
Warm-up phase (function check activated)
Enclosure
Weight
Degree of protection
Electrical characteristics
Electrical safety
EMC interference immunity
(with safety extra-low voltage (SELV) with safe
isolation)
Auxiliary power supply
Power consumption
Fuse values
Electrical inputs and outputs
Series 6 and ULTRAMAT 23
Compact Operating Instructions, 01/2024, A5E45779144002-AB
Maximum of 4
2 per component, freely adjustable between smallest and largest measur‐
ing range
Linearized
LCD with LED backlighting and contrast control, 80 characters (4 lines à
20 characters); function keys
Front panel vertical
No vibrations
Maximum vibration 5 m/s
2
30 minutes if no UV module is installed
60 minutes if UV module is installed
Approx. 10 kg (22 lbs.) The weight varies according to the ordered variant.
IP40 in accordance with EN / IEC 60529 for the versions 7MB235x (TÜV)
IP20 in accordance with EN / IEC 60529 for the versions 7MB233x (standard)
 
In accordance with EN / IEC 61010-1,
overvoltage category II
In accordance with standard requirements of NAMUR NE21 or
EN IEC 61326-1
100 V AC, +10%/-15%, 50 Hz,
120 V AC, +10%/-15%, 50 Hz,
200 V AC, +10%/-15%, 50 Hz,
230 V AC, +10%/-15%, 50 Hz,
100 V AC, +10%/-15%, 60 Hz,
120 V AC, +10%/-15%, 60 Hz,
230 V AC, +10%/-15%, 60 Hz
Max. 60 VA depending on ordered version
200 V/230 V AC; T 630 mA/L 250 V
see inscription on rear of device
100 V/120 V AC; T 1.25 A/L 250 V
see inscription on rear of device
7
51

Advertisement

Table of Contents
loading

Table of Contents