Troubleshooting - Savaria Track Bridge Owner's Manual

Ceiling lift
Table of Contents

Advertisement

7. TROUBLESHOOTING

Do NOT open the device casing. Only an Authorized
Savaria Dealer is qualified to open it. Modifications made
to the device by someone other than a qualified
technician may cause serious injury.
For complementary informations, please refer to Device Controls section.
Problem
The unit starts and stops
repeatedly.
The track bridge doesn't
beep or engage.
The track bridge emits a beep
every minute.
Ceiling lift stuck in the track
bridge halfway engaged.
Part No. 001300-EN, 10-m10-2023 Track Bridge /Track Bridge Bariatric Owner's Manual
WARNING
Action
Verify the connection from the power outlet to the
power source, and then to the device.
If the problem persists and the device still does not
operate, contact your Authorized Savaria Dealer for
service.
Make sure the power switch is in the ON position.
Make sure the ceiling lift is equipped with a
magnetic block and is inside the track bridge. Make
sure the track bridge sections are aligned or that the
mobile track speed isn't to fast when aligning with
the fixed track. Ensure that the track bridge is
properly energized, see following.
Verify the power outlet is energized, the charger is
properly plugged into the power outlet and
connected to the CCR strip. Verify also that the CCR
brushes are plugged into the track bridge.
Use the manual operation (electrical or
mechanical). See Section 5.