8. ELECTROMAGNETIC COMPATIBILITY

Electromagnetic Compliance

The device has been tested for compliance with current regulatory standards in
regards to EMI (electromagnetic interference).
Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If
such use is necessary, this equipment and the other equipment should
be observed to verify that they are operating normally.
Use of accessories and cables other than those specified or provided by
the manufacturer of this equipment, could result in increased
electromagnetic emissions or decreased electromagnetic immunity of
this equipment, and result in improper operation.
Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than
30 cm (12 inches) to any part of the device, including cables specified
by the manufacturer. Otherwise, degradation of the performance of
this equipment could result.
Electromagnetic Emissions - For All Equipment and Systems
The device is intended for use in the electromagnetic environment specified below.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Track Bridge /Track Bridge Bariatric Owner's Manual Part No. 001300-EN, 10-m10-2023
Guidance and Manufacturer's Declaration -
Compliance
Group 1
Class B
Not applicable
Not applicable
WARNING
Electromagnetic environment - guidance
The device uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference with
nearby equipment
The device is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
supplying buildings used for domestic purposes.