Test 014: Adjusting and Displaying the Flow
This test allows you to adjust the flow in steps of 100 mL. The measured flow is continually
displayed.
The l:E valve is switched on at the beginning of the test and switched off again at the end of
the test.
Test Setup
flowmeter
A
Important!
In the Air Mix mode, generate a backpressure (e.g. using a hose
clamp) in order to supply the appropriate flow.
• Adjust the mode switch to IPPV.
• Set the selector switch for PEEP/CPAP to 0 mbar.
• Select the Extended Service Mode (ESM).
Press the Reset key or the Info key to select Test 014.
Press the Info key and the Reset key simultaneously to activate Test 014.
DS-Modus 5503.160 Oxylog 2000
11/97
Service-Mode SW 2.xx
Page 24