Drager Medizintechnik
Drager
3.4
Test 014: Adjusting and Displaying the Flow
This test allows you to adjust the flow in steps of 100 mL. The measured flow is continually
displayed.
The l:E valve is switched on at the beginning of the test and switched off again at the end of
the test.
Test Setup
flowmeter
A
Important!
In the Air Mix mode, generate a backpressure (e.g. using a hose
clamp) in order to supply the appropriate flow.
• Adjust the mode switch to IPPV.
• Set the selector switch for PEEP/CPAP to 0 mbar.
• Select the Extended Service Mode (ESM).
•
Press the Reset key or the Info key to select Test 014.
•
Press the Info key and the Reset key simultaneously to activate Test 014.
DS-Modus 5503.160 Oxylog 2000
11/97
Service-Mode SW 2.xx
Page 24