Drager Medizintechnik
Important!
If the values shown in the flow setiings table are not reached, then carry out Test 012 and
Test 013.
Press the Info key and the Reset key simultaneously to finish Test 014.
The Oxylog 2000 automatically returns to the Extended Service Mode (ESM).
Exiting the Extended Service Mode (ESM).
• Switch the Oxylog 2000 off and on again to return to the Oxylog 2000 normal operating
mode.
DS-Modus 5503.160 Oxylog 2000
11/97
Service-Mode SW 2.xx
Page 26