3.2.2
Software Version 3.xx
Display:
Set Pmax = 60 mbar.
0 mbar pressure
• Set the mode switch for Pmax to 60 mbar.
Press the Info key and the Reset key simultaneously to start the calibration mode.
NoAirMix
xx L/min
ENTER = calibration point
Press the Info key or the Reset key to set the flow so that the value shown on the display
screen matches the value measured on the external flowmeter.
Press the Info key and the Reset key simultaneously to save the following calibration values
in the EEPROM.
Calibration Value
Tolerances for Set Flow Values
10
9.5-10.5 L/min
20
19.0-21.0 L/min
30
29.0-30.0 L/min
50
49.0-51.0 L/min
60
59.0 - 60.0 L/min
The last calibration value is calibrated. Oxylog 2000 automatically returns to the Extended
Service Mode (ESM).
Exiting the Extended Service Mode (ESM)
If it is switched off and then on again, Oxylog 2000 returns to normal operation.