CURRENT REQUIREMENT – US System
Precision RXi
COMPONENT
Generator – 65kW
Generator – 80kW
Positioner
Digital Transformer
OTS
CURRENT REQUIREMENT – European System
Precision RXi
COMPONENT
Generator – 65kW
Generator – 80kW
Positioner
Digital Transformer
65 kW GENERATOR POWER DEMAND – US Systems
Continuous power demand
Minimum recommended distribution transformer rating
Maximum momentary mA
Maximum momentary kVp
65 kW GENERATOR POWER DEMAND – European Systems
Continuous power demand
Minimum recommended distribution transformer rating
Maximum momentary mA
Maximum momentary kVp
80 kW GENERATOR POWER DEMAND – US Systems
Continuous power demand
Minimum recommended distribution transformer rating
Maximum momentary mA
Maximum momentary kVp
80 kW GENERATOR POWER DEMAND – European Systems
Continuous power demand
Minimum recommended distribution transformer rating
Maximum momentary mA
Maximum momentary kVp
Page 72 / 111
12
EVISION
CURRENT
MOMENTARY
105A
130A
N/A
N/A
N/A
CURRENT
MOMENTARY
125A
155A
N/A
N/A
P
RECISION
CONTINUOUS
5A(standby current)
5A(standby current)
17.4A
4.4 A
1.3A
CONTINUOUS
5A(standby current)
5A(standby current)
17.4 A
4.4 A
2.4 kVA at 480V, 50/60 Hz
85 kVA at 480V, 50/60 Hz
800mA at 480V, 50/60 Hz
120kVp at 480V, 50/60 Hz
2.0 kVA at 400V, 50/60 Hz
85 kVA at 400V, 50/60 Hz
800mA at 400V, 50/60 Hz
120kVp at 400V, 50/60 Hz
2.4 kVA at 480V, 50/60 Hz
105 kVA at 480V, 50/60 Hz
1000mA at 480V, 50/60 Hz
120kVp at 480V, 50/60 Hz
2.0 kVA at 400V, 50/60 Hz
105 kVA at 400V, 50/60 Hz
1000mA at 400V, 50/60 Hz
120kVp at 400V, 50/60 Hz
June 2009
RX
S
P
-I
I
YSTEM
RE
NSTALLATION