5 Commissioning and First Operating Steps | 5.5 Modifying Gun Parameters

5.4.8 Saving the Image

Procedure
1. In the SEM Controls panel, select the Scanning tab.
2. Select Freeze on = End Frame from the drop-
3. Click Freeze.
4. From the Menu Bar, select File > Save Image.
5. To change the save path, click Change Directory.
6. To confirm the selected path, click Select Folder.
7. Enter the filename in the Filename input field.
8. Click Save <file name>.tif.
9. To continue imaging, click Unfreeze in the Scanning tab.

5.5 Modifying Gun Parameters

5.5.1 Selecting the Gun Mode

The microscope can operate in different gun modes:
§
§
Imaging Gun Mode In Imaging gun mode, the temperature of the Schottky emitter and the extraction voltage are re-
duced in comparison to the Analytic gun mode. This leads to a reduction of the energy spread of
the primary electrons.
Imaging gun mode is suitable for lower probe currents. Imaging gun mode is especially useful at
low kV to reduce chromatic aberration and to achieve a better resolution.
96
down list.
à
A red dot at the right bottom of the Image Area indicates that the image is frozen.
à
The Export TIFF dialog is displayed.
à
A file explorer window is displayed.
Imaging