Saving The Image; Modifying Gun Parameters And Optibeam Mode; Using The Automatic Gun Alignment Functions - Zeiss EVO Instruction Manual

Scanning electron microscope
Table of Contents

Advertisement

ZEISS
5. In the Stigmation navigation box, use the
6. To deactivate the reduced raster, in the Toolbar, click the REDUCE icon.
5.4.10.7 Adjusting Final Magnification, Contrast and Brightness
Procedure
1. Reduce the magnification to the required value (e.g. 25,000 x).
2. To activate auto contrast and brightness, click the LEVELS icon in the Toolbar.

5.4.11 Saving the Image

Procedure
1. Once a steady contrast and brightness level is reached, click the PHOTO icon in the Tool-
2. From the Menu Bar, select File > Save Image.
3. To change the save path, click Change Directory.
4. To confirm the selected path, click Select Folder.
5. Enter the file name in the Filename input field.
6. Click Save <file name>.tif.
7. To continue imaging, middle-click the PHOTO icon.

5.5 Modifying Gun Parameters and Optibeam Mode

5.5.1 Using the Automatic Gun Alignment Functions

Automatic gun align is used to automatically align the shift and/or the tilt of the beam, to the cen-
ter of the column.
Two automatic gun align functions are available:
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006
scroll bars or the red marker to obtain the
sharpest possible image.
INFO:
The specimen detail should just be pul-
sating without shifting.
INFO:
To obtain optimum results, alternately
correct focus and astigmatism.
bar.
INFO:
This runs a macro that automatically changes the scan speed to 8, the noise reduc-
tion to line integration with 4 lines and freezes the image at the end of the frame.
à
A red dot at the right bottom of the image area indicates that the image is frozen.
à
The Export TIFF dialog is displayed.
à
A file explorer window is displayed.
5 First Operating Steps | 5.5 Modifying Gun Parameters and Optibeam Mode
79