Alarm Function - UNI-T UTE9800+ Series User Manual

Smart digital power meter
Table of Contents

Advertisement

User's Manual
Explanation
The unit of alarm delay is S,the range can set to 0~99.9.

5.3 Alarm Function

5.3.1 Turn on/off Alarm Function of UTE9802+/UTE9811+
When the upper/lower limit is different but as "0", it represents the alarm function is enabled.
When the upper/lower limit is "0"at the same time, it represents the alarm function is forbidden.
Alarm Function
After the alarm function is enabled, the system waits to connect the load. When the system detects the voltage
and current are both greater than "0", it will recognize the load is connected and start to count the time. After
the alarm delay, the system will compare the measured value with the upper/lower limit, if the measured value
is within the upper/lower limit, the interface displays "OK"; if the measured value is higher than the upper limit,
the interface displays "Hi", and the beeper will alarm; if the measured value is lower than the lower limit, the
interface displays "Lo", and the beeper will alarm. When the load is removed, the system waits for load access
again, the alarm sign is eliminated, and the beeper is turned off.
5.3.2 Turn on/off Alarm Function of UTE9806+
·Steps
1. Press【Shift】and【Setup】key to enter Utility menu;
2. Press【▲】or【▼】key to select ALARM, as shown in the following figure;
3. Press 【 OK 】 key to enter ALARM menu , press 【 ▲ 】 or 【 ▼ 】 key to select ON or OFF ;
Instruments.uni-trend.com
UTE9800+ Series Smart Digital Power Meter