Instructions for use and disinfection of
tonometer measuring prisms
NOTE
319300-9231-000-AddGA01-GB-240320
2.6 List of disinfectants with good material compatibility with ZEISS measuring
2.6 List of disinfectants with good material
compatibility with ZEISS measuring prisms
The disinfectants listed below were only tested on their material
compatibility with ZEISS measuring prisms.
Damage due to improper cleaning and disinfection
The effectiveness of the disinfectants has not been tested for
correct disinfection of the tonometer measuring prisms.
The sole purpose of the list below is to provide the user with a
choice of suitable disinfectants which may be used for the
validation of their own disinfection process.
ZEISS rejects all liability if any of the disinfectants mentioned in the
list below is used without a proper validated cleaning and disin-
fection process.
The disinfectants may only be used with the concentration
mentioned on the list below. Higher concentrations can damage
the measuring prisms. In this case ZEISS rejects all liability.
ZEISS rejects all liability if a disinfectant is used which does not
appear on the list below.
The efficacy of the disinfectants on ZEISS tonometer measuring
u
prisms must be guaranteed by the user or the responsible
person for reprocessing by validating their own disinfection
process.
2 Safety notes
prisms
9 / 18