Bindicator PRD1000 Instruction Manual page 65

Analog input process meter
Table of Contents

Advertisement

PRD1000 Analog Input Process Meter
Display
Parameter
break
Break
Maximum
mAx
Minimum
min
4 mA output
4 mA
20 mA
20 mA output
user
User I/O
F1 function key
F1
F2 function key
F2
F3 function key
F3
F4 function
F4
ICal
Internal source
calibration
Current
C CAL
calibration
Current low
C lo
Current high
C Hi
Voltage
v CAL
calibration
v Lo
Voltage low
v Hi
Voltage high
Action/Setting
Set input break condition operation
Program maximum mA output allowed
Program minimum mA output allowed
Enter mA output value read by milliamp
meter with at least 0.001 mA resolution
Enter mA output value read by milliamp
meter with at least 0.001 mA resolution
Assign function keys and digital I/O
Assign F1 function key
Assign F2 function key
Assign F3 function key
Assign F4 function (digital input)
Enter internal source calibration (used for
scaling the meter without a signal source)
Calibrate 4-20 mA current input (internal
reference source used for scaling the input)
Calibrate low current input (e.g. 4 mA)
Calibrate high current input (e.g. 20 mA)
Calibrate voltage input
Calibrate low voltage input (e.g. 0 V)
Calibrate high voltage input (e.g. 10 V)