MRC DFO-600 Operation Manual

MRC DFO-600 Operation Manual

Forced convection oven

Advertisement

Quick Links

Operation Manual
Forced Convection Oven
DFO-600/DFO-720/DFO-1008/DFO-2160
1
3, Hagavish st. , Holon,Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529
mrc@mrclab.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DFO-600 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for MRC DFO-600

  • Page 1 Operation Manual Forced Convection Oven DFO-600/DFO-720/DFO-1008/DFO-2160 3, Hagavish st. , Holon,Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com...
  • Page 2: Table Of Contents

    3.2 Setting Timer 4. Setting parameter 4.1 Lockout 4.2 User Calibration 4.3 Auto-Tuning 5. Error code for temperature controller 6. Troubleshooting Guide 7. General Maintenance 8. Warranty 3, Hagavish st. , Holon,Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com...
  • Page 3 1. Specifications MODEL DFO-600 DFO-720 DFO-1008 DFO-2160 SPECIFICATIONS Ambient +40℃ to 250℃ TEMP.RANGE ±0.1℃ (LED Display) Uniform ±1.5~3℃ TEMP.ACCURACY TEMP.CONTROL PID Controller TEMP.DISPLAY LED Digital Display TEMP.SENSOR K TYPE SAFETY DEVICE EGO (DOUBLE TEMPERATURE PROTECTION) TIMER 1 minute to 4553.6 minutes or continuous...
  • Page 4: Panel & Key Description

    LAMP If the timer reaches the end, the lamp will be on, and the whole heating system will stop output 3, Hagavish st. , Holon,Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com...
  • Page 5: Operation

    Light on when heater is activated ⑩ MANU Lamp flashes during Manual Mode operates ⑪ COMM Communication indicator ⑫ TUNE Lamp flashes during auto-tuning operates ⑬ Lock indicator 3, Hagavish st. , Holon,Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com...
  • Page 6: Setting The Temperature

    (display temperature), all settings have been completed When actual temperature reaches the setting temperature, timer will start counting. When time out. Timer indicator LED lights. 3, Hagavish st. , Holon,Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com...
  • Page 7: Setting Parameter

    Set the PASS value to your desired function PASS Value Access Rights = 1000 All parameters are changeable = 1008 Only user menu parameters changeable (original value) 3, Hagavish st. , Holon,Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com...
  • Page 8: User Calibration

    PID control by using its new PID values. The PID values obtained are stored in nonvolatile memory. Note: To enable auto-calculating the lock parameter should be set with 1000. 3, Hagavish st. , Holon,Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com...
  • Page 9: Error Code For Temperature Controller

    1 mA if 4-20 mA is used, or input voltage below 0.25V if 1 - 5V is used AADER A to D converter or related Return to factory for repair. component(s) malfunction 3, Hagavish st. , Holon,Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com...
  • Page 10: Troubleshooting Guide

    Operate at higher temperature, or temperature. turn on the compressor. Temperature controller is defective. Replace. Circulating fan doesn’t turn. Replace fan motor. Controller needs auto- tuning. Perform auto-tuning. 3, Hagavish st. , Holon,Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com...
  • Page 11: General Maintenance

    5. This warranty card is invalid without company stamp, and will not be re-issued in case it is lost. 8-2 Warranty Card Product Name Model Number Serial number Date of purchase Warranty period Company chop (Invalid without company stamp) 3, Hagavish st. , Holon,Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com...