Thank you for choosing this Steam Sterilizer. TABLE OF CONTENTS Please read the instructions manual carefully in order to install the 1.GENERAL ------------------------------------------------------------ 1 equipment and operate the equipment in an efficient manner. 2.TECHNICAL PARAMETER ---------------------------------------- 2 3.PACKING CONTENT ----------------------------------------------- 3 4.INSTALLATION ---------------------------------------------------- 4 5.OPERATION ---------------------------------------------------------- 5 5.1 READY ------------------------------------------------------------- 5...
1 General 2 Technical Parameter This sterilizer described in this manual is intended for the sterilization of re-useable surgical instruments and material. Item It operates automatically with 134 ℃ and 121 ℃ sterilization temperatures. It has been produced in accordance with the EN 13060. Φ200mmX360mm Φ230mmX360mm Chamber...
4 Installation 3 Packing Content * There must leaves 10cm gap around sterilizer, and 20cm on top side. the Item Quantity clearance required for the movement of the door : leave an at least Steam sterilizer 350mm fan-shaped space in front of the door. Instrument tray * The place which sterilizer located must be ventilated, make sure that the Instrument tray rack...
5.1.4 Filling the distilled water 5 Operation Open the top lid, and fill the tank with distilled water by cup or tank. 5.1 Ready When you hear a beep signal, it means the water level exceeds the 5.1.1 Open the door and take out all the instrument tray and other max.
* Turn the sterilization paper in order to set the plastic part downward (tray side) and the paper part upward. SOLID WRAPPED Always wrap the tools in case of prolonged store. TEXTILE OR DRYING DRYING 5.3 Selecting the sterilization program 5.3.1 LCD LIQUID It displays the cycle temperature,...
Page 7
5.4.3 Sterilization cycle completion 5.4.2 Start the sterilization program. Press START button the machine will begin the cycle automatically. After a cycle is completed, the printer will start work and print the report of the sterilization cycle data.(if you connect the printer) It will take 30-75 minutes.
5.5 Test programs (CLASS B) 6 Advanced Setting 5.5.1 Press PROGRAM button, select the "B&D TEST". 6.1 Enter the setting 5.5.1.1 Put the Bowie-Dick pack into the 6.1.1 Holding the START button and open the chamber. Then close the door and main switch.
Page 9
6.6 USB Flash memory (Optional) 6.4 S3 state 6.4.1 Adjust the holding time of sterilization and drying time. Insert the USB stick to the slot Press PROGRAM button to select the program.( The information will output to USB stick after the cycle finish Press TEMP button to select the temperature of program.
7 Maintenance 7.4 Clean the door seal 7 . Mainte n ance Clean the door seal weekly, with non-plush cloth saturated with the distilled water. Frequency Operation Cleaning the door seal Daily Cleaning the external surface Cleaning the reservoir Weekly Cleaning the chamber Every 3/6 monthly Replacing the bacteriological filter...
Page 12
7.6 Replacement of the door seal 7.7 The drain valve 7.6.1 Fully open the door. 7.6.2 Remove the door seal ring carefully by hand. 7.6.3 Clean the door seal ring carefully with a non-plush cloth saturated with distilled water. 7.6.4 Moisten the new seal ring with medical disinfectant . 2.Set the drain valve to the open 7.6.5 Insert the new seal ring and press in sequence as the 1.Push the hose to the drain valve...
9 Error codes 10 Safety devices (1)Main fuses Protection of the whole equipment against possible failures of the Code Description Proposed solution heating resistor . Steam generator temperature Check the steam generator temperature Action: Interruption of the electric power supply. sensor error sensor (2)Thermal cutouts on the mains transformer windings...
APPENDIX 2 APPENDIX 1 Characteristics of the feeding water DIAGRAMS OF THE STERILIZATION PROGRAMS PROGRAM TYPE 。 FEED WATER CLASS B CONDENSATE DESCRIPTION ≤10 mg/l ≤1.0 mg/kg Evaporate residue 14 30 Unwrapped solid 4 00 1 20 SOLID material ≤0.1 mg/kg Silicium oxide sio ≤1 mg/l 30 45...
APPENDIX 3 ELECTRICAL DRAWING ELECTRICAL DRAWING CLASS N CLASS B USB DATA USB DATA OUTPUT OUTPUT THERMAL THERMAL PROTECTOR PROTECTOR CHAMBER CHAMBER HEATER HEATER STEAM STEAM HEATER HEATER THERMAL THERMAL PROTECTOR PROTECTOR WATER WATER PUMP PUMP VACUUM VACUUM PUMP PUMP ~...
APPENDIX 4 HYDRAULIC DRAWING HYDRAULIC DRAWING CLASS B CLASS N Air filter Safety valve Safety valve Pressure Pressure sensor sensor Water pump Pump Chamber Chamber Condenser Condenser T e st P o r t Distilled water tank Distilled water tank Vacuum pump Used water tank Used water tank...