PSVPro™ Ventilation Mode
Pressure supported ventilation with apnea backup
Clinician sets the Psupport and PEEP levels, the patient establishes the rate, inspiratory flow and inspiratory time
Tidal volume is determined by the pressure, lung characteristics and patient effort
Apnea backup mode (SIMV PCV) is provided if the patient stops breathing
In backup mode, the Backup Mode active alarm is shown until PSVPro™ Ventilation Mode is reinstated or another
ventilation mode is selected
PSVPro mode settings:
Psupport
PEEP
Trig Window
Flow Trigger
End of Breath
Pmax
Backup Time
Pinsp
RR
Tinsp
Rise Rate
Exit Backup

CPAP + PSV mode

Provides a constant support pressure for spontaneously breathing patients once the ventilator senses an
inspiratory effort
The patient establishes the rate, inspiratory flow, and inspiratory time
The tidal volume is determined by the pressure, lung characteristics, and patient effort.
Spontaneous breaths that occur are indicated by a color change in the waveform
CPAP + PSV settings:
Psupport
PEEP
Flow Trigger
End of Breath
Pmax
Pinsp
Minimum RR
Tinsp
Rise Rate
T
Peep
insp
T
insp
Peep
Time
34
Time

Hide quick links: