Additional alarm settings
The following alarms can be turned off to reduce
nuisance alarms during manual ventilation if clinically
appropriate:
1. CO
Alarms: Use the CO
2
the CO
'Apnea' alarm, EtCO
2
FiCO
high and Absorbent OK? alarms during
2
manual ventilation.
2. Vol Apnea Alarm: Use the Vol Apnea Alarm setting
to turn off the volume apnea alarm during manual
ventilation.
3. MV/TV Alarms: Use the MV/TV Alarms setting to
turn off the MV and TV alarms. If MV/TV Alarms is
set to Off during manual ventilation, the alarms
remain off until the Bag/Vent switch is set to Vent or
the MV/TV Alarms is set to On. If the MV/TV
Alarms is set to Off during mechanical ventilation,
the alarms remain off when manual ventilation
starts.
Note: CO
Alarms and Apnea Alarm selection will not
2
be visible unless enabled by a Super User. CO
and Vol Apnea Alarm can only be turned off while in
manual ventilation.
Alarms setting to turn off
2
low, EtCO
high, and
2
2
Alarms
2
1
28
2
3