Environmental Specifications - Keysight U1832A User Manual

Usb noise sources
Table of Contents

Advertisement

Environmental Specifications

U1832A/B/C and U1833A/B/C/D USB Noise Sources are designed for indoor use
in an area with low condensation.
specifications for these instruments.
Samples of this product have been type tested in accordance with the Keysight
NOTE
Environmental Test Manual and verified to be robust against the Environmental
stresses of Storage, Transportation and End-use; those stresses include, but are
not limited to temperature, humidity, shock, vibration and altitude conditions.
The primary source of Keysight Environmental Test standards is the International
Electro-technical Commission (IEC) standards.
Table 3-2
Environmental condition
Temperature
Humidity
Shock
Vibration
Keysight U1832A/B/C and U1833A/B/C/D User's Guide
Table 3-2
U1832x/U1833x Environmental Specifications
Requirement
Operating condition
– 0 °C to 55 °C
Storage condition
– –40 °C to 70 °C
Operating condition
– Up to 95% RH at 40 °C (non-condensing)
Method 1: 23°C/50% RH to 40°C/95% RH
End-use handling shock
– ΔV: 1.6 m/s (60 in/s) ±5%, duration <3 ms
Transportation shock
– 50 g, ΔV: 8 m/s ±10%
Operating random vibration
– 5-500 Hz, 0.21 Grms
Survival Swept Sine Vibration
– 5 Hz-500 Hz-5 Hz, 0.5 Grms
Survival random vibration
– 5-500 Hz, 2.09 Grms
Packaging Drop
– 48" drop height
Specifications
shows the environmental
3
37

Advertisement

Table of Contents
loading

This manual is also suitable for:

U1832bU1832cU1833aU1833bU1833cU1833d

Table of Contents