Immunity Test
Electrical fast
transient/burst
EC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. main voltage prior to application of the test level.
12.5 GUIDANCE AND MANUFACTURER'S DECLARATION –
ELECTROMAGNETIC EMISSIONS
The concentrator is intended for use in home, institution, vehicle and other transportation and mobile envi-
ronments. The user of the concentrator should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage fluctuations /flicker
emissions
IEC 61000-3-3
ELECTRICAL ISOLATION DEVICE
The external power supply provides the means for electrical isolation where the AC inlet is incorporated
into the power supply.
©2023 Inogen, Inc. All rights reserved.
IEC 60601 Test Level
± 2 kV for power
supply lines
± 1 kV line(s) to
line(s)
0% UT for 0.5 cycle at
0 °, 45 °, 90 °, 135 °,
180 °, 225 °, 270 °, and
315 °.
0% UT for 1 cycle
70% UT for 25/30 cycle
0% UT for 200/300 cycle
30 A/m
Compliance
Group 1
Class B
Class A
Complies
Electromagnetic Environment Guidance
Mains power quality should be that of a typical home,
institution, vehicle or other transpiration and mobile
environments.
Mains power quality should be that of a typical home,
institution, vehicle or other transpiration and mobile
environments.
Mains power quality should be that of a typical
home, institution, vehicle and other transportation
and mobile environments. If the user of the Rove 6
requires continued operation during power mains
interruptions, it is recommended that the device be
powered from an uninterrupted power supply.
Power frequency magnetic fields should be at levels
characteristic of a typical home, institution, vehicle
and various mobile environments. Power frequency
magnetic fields from common appliances in the home
are not expected to affect
the device.
Electromagnetic Environment
Guidance
The concentrator uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and not likely to cause any
interference in nearby equipment.
The concentrator is suitable for use in
all establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Page 35 of 40