Inogen One G3 Technical Manual
Hide thumbs Also See for One G3:

Advertisement

INOGEN ONE G3 OXYGEN CONCENTRATOR
TECHNICAL MANUAL
96-03996-00-01
Revision F
This Manual applies to the following Inogen, Inc. products:
 Inogen One G3 Oxygen Concentrator, model # IO-300
 Inogen One AC Power supply, model # BA-301
 Battery, model # BA-300
 Battery, model # BA-316
CONFIDENTIAL and
INOGEN ONE G3 OXYGEN CONCENTRATOR
Page 1 of 17
96-03996-00-01 revF
DCR-15-284
PROPRIETARY
TECHNICAL MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the One G3 and is the answer not in the manual?

Questions and answers

Janet H Marino
May 18, 2025

How to install new cylinders in inogen one

Subscribe to Our Youtube Channel

Summary of Contents for Inogen One G3

  • Page 1 96-03996-00-01 Revision F This Manual applies to the following Inogen, Inc. products:  Inogen One G3 Oxygen Concentrator, model # IO-300  Inogen One AC Power supply, model # BA-301  Battery, model # BA-300  Battery, model # BA-316...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS     INFORMATION FOR PROVIDERS OF THE INOGEN ONE G3 ................ 3     ......................... 3 1.1. AUTION AND ARNING TATEMENTS     SETTING UP A PATIENT ON INOGEN ONE G3 ..................... 5     ..........................5 2.1.
  • Page 3: Information For Providers Of The Inogen One G3

    Inogen One G3 and its accessories. Proper care in relaying this information will not only enhance the user’s experience with the Inogen One G3, but will also protect the patient, prolong the life of device, help avoid...
  • Page 4 Inogen One G3. In many cases, warnings and cautions have been included in the Inogen One G3 User Manual. CONFIDENTIAL and INOGEN ONE G3 OXYGEN CONCENTRATOR Page 4 of 17 96-03996-00-01 revF...
  • Page 5: Setting Up A Patient On Inogen One G3

    2. SETTING UP A PATIENT ON INOGEN ONE G3 2.1. Indications for Use The Inogen One G3 Oxygen Concentrator is used on a prescriptive basis by patients requiring supplemental oxygen. It supplies a high concentration of oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient.
  • Page 6: Selecting The Proper Flow Setting

    4. Plug into closest AC outlet 5. Turn on the Inogen One G3 by pressing the on/off button. 6. Set the Inogen One G3 to correct flow setting prescribed by the physician or clinician by pressing the + (increase flow) or –...
  • Page 7: Servicing The Inogen One G3

    Inogen suggests that each patient be titrated (a) while sedentary, (b) while active or ambulatory. 2.4.4. Use with the Inogen Satellite Conserver The Inogen One G3 is not compatible with the Inogen One Satellite Conserver 3. SERVICING THE INOGEN ONE G3 3.1.
  • Page 8 The concentrator’s display will show the hour meter, the unit’s serial number, and the software version installed on the device. Do not disassemble the Inogen One G3 or any of the accessories or attempt any maintenance other than tasks described in this Technical Manual unless you have completed a training course through Inogen.
  • Page 9 Inogen suggests that this filter be replaced between patients. To replace the Product Filter: 1. Use the Cannula Barb Tool (available from Inogen as RP- 102) to access the product filter. The tool has two prongs which mate with two indentations located on the surface of the metal cannula barb fitting on the Inogen One 2.
  • Page 10  Charging Status 3.1.8. Data Download To download data from the Inogen One G3, you will need a USB mass storage device. If one of the following drives cannot be obtained, the drive must be less than 2GB in capacity and have a native file format of FAT32.
  • Page 11: Maintenance By The Patient

     Transcend JetFlash V30 2GB PN:TS2GJFV30  Patriot Signature 2GB PN: PSF2GUSB To collect data: a) With the Inogen One G3 off, insert the USB mass storage device into the USB port behind the blue rubber cover b) Plug in external power to the concentrator c) Press the “alarm bell”...
  • Page 12 3.2.4. Battery Care and Maintenance The Inogen One G3 Lithium Ion Battery requires special care to ensure proper performance and long life. Use only Inogen One G3 Batteries with the Inogen One G3 Oxygen Concentrator.
  • Page 13 This power draw will empty a full battery in approximately 2-3 weeks. Leaving a Battery in the Inogen One G3 while the device is unused will irreversibly damage the battery. After such storage, the battery may not be able to recharge or its CAUTION life cycle and/or capacity will be greatly diminished.
  • Page 14: Expected Service Requirements

    Sieve Valves 25,000 run hours Intake 20,000 run hours (without smoke or other abnormal contaminants) Filter Cooling Fan 25,000 run hours 4. INOGEN ONE G3 SYSTEM SPECIFICATIONS INOGEN ONE G3 CONCENTRATOR Length Width Height No battery 22.2 8.75 18.4 7.25 Dimensions: 22.2...
  • Page 15: Inogen One G3 Error Code Table

    Ingress of Water: w/carry bag: IP 22 Degree of Safety for Application Not suitable for such in Presence of Anesthetic Gases: applications 5. INOGEN ONE G3 ERROR CODE TABLE Error Code Explanation Possible Causes & Number Troubleshooting Instructions Low voltage on power...
  • Page 16 A Breath is detected but Check for kinked tubing and Error oxygen is not being listen for proper exhaust delivered noises. If problem CONFIDENTIAL and INOGEN ONE G3 OXYGEN CONCENTRATOR Page 16 of 17 96-03996-00-01 revF DCR-15-284 PROPRIETARY TECHNICAL MANUAL...
  • Page 17: Error Recall

    Error 1. 7. CONTACTS FOR MORE INFORMATION Inogen Website: Patient Issues: http://www.inogenone.com Corporate: http://www.inogen.net Inogen : 326 Bollay Dr Goleta, CA 93117 1-866-765-2800 clientservices@inogen.net CONFIDENTIAL and INOGEN ONE G3 OXYGEN CONCENTRATOR Page 17 of 17 96-03996-00-01 revF DCR-15-284 PROPRIETARY TECHNICAL MANUAL...

This manual is also suitable for:

Io-300