7.5.3 LOW PRIORITY ALARM (LEVEL 2)
The following low priority alarms are accompanied by one beep and a solid yellow light.
Display Icon
7.5.4 LOW PRIORITY ALARM (LEVEL 3)
The following low priority alarms are accompanied by two beeps and a solid yellow light.
Display Icon
©2023 Inogen, Inc. All rights reserved.
Description
Remove battery to cool:
Remove battery to cool.
Check battery: Check battery.
Description
Replace columns: Column
replacement is required within
30 days.
Extended start up: Oxygen
concentration is <87% two
minutes after the device's start
up sequence and at least 10
breaths have been detected
within the last minute.
Description
Battery low, attach plug:
Battery power is low with less
than 10 minutes remaining.
Page 22 of 40
What To Do
The battery needs to be removed and
must be cooled before reuse.
Check the connection of your battery
and ensure that it is properly attached
and latched to the concentrator. If
the battery error persists with same
battery, stop using the battery and
switch to a new battery or remove the
battery and operate the concentrator
using an external power supply.
What To Do
Contact your equipment
provider to arrange for service
and/or order new columns from the
manufacturer.
Wait a few minutes to see if the
oxygen concentration improves
(alarm will clear). If condition
persists, a secondary alarm will
sound. Follow the instructions for
that alarm or contact your
equipment provider. If alarm occurs
frequently at start up, this may
indicate that maintenance (column
replacement) will soon be required.
What To Do
Attach an external power supply turn
off and insert a fully charged battery.
96-13115-00-01 B