Inogen Rove 6 User Manual page 23

Hide thumbs Also See for Rove 6:
Table of Contents

Advertisement

Display Icon
7.5.5 MEDIUM PRIORITY ALARMS (LEVEL 4)
The following medium priority alerts are accompanied by three beeps, repeated every 25 seconds,
and a flashing yellow light.
Display Icon
©2023 Inogen, Inc. All rights reserved.
Description
Oxygen low: The concentrator
has been producing oxygen at
a slightly low level (≤82%) for a
period of 10 minutes.
Service soon: The
concentrator requires servicing
at the earliest convenience.
The concentrator is operating
to specification and may
continue to be used.
Battery HOT warning: The
battery temperature is nearing
the temperature limit while
concentrator is running on
battery power.
System HOT warning:
Concentrator temperature is
nearing temperature limit.
Description
No-breath-detect: check
cannula: The concentrator
has not detected a breath for
60 seconds.
Oxygen error: Oxygen output
concentration has been below
50% for 10 minutes.
Page 23 of 40
What To Do
If condition persists, contact your
equipment provider.
Contact your equipment provider to
arrange for service.
If possible, move the concentrator to
a cooler location or power unit with
an external power supply and remove
battery. If condition persists, contact
your equipment provider.
If possible, move the concentrator to a
cooler location. Ensure air intake and
outlet vents have clear access and
particle filters are clean. If condition
persists, contact your equipment
provider.
What To Do
Check that cannula is connected to
concentrator, there are no kinks
in tubing and the cannula is
positioned properly in your nose.
If condition persists, switch to your
backup oxygen source and contact
your equipment provider to arrange
for service.