10.3 FILTER CLEANING & REPLACEMENT (RP-501)
The particle filters must be cleaned weekly to ensure the ease of air flow.
To clean:
1. Remove the battery from the device.
2. Remove the particle filters from both intake ends of the device.
3. Clean the particle filters with a mild liquid detergent and water, rinse in water and dry fully before reuse.
To purchase additional particle filters, contact your equipment provider or the manufacturer Inogen, at
Inogen.com or by calling 1-877-466-4364.
10.4 CANNULA BARB OUTPUT FILTER REPLACEMENT (RP-506)
The cannula barb connects the gas pathway to the cannula while the output filter is designed to protect
the user from breathing in small particles when using the device. The output filter is located behind the
cannula barb and should be replaced between patients or when replacing the cannula barb. To replace the
cannula barb and output filter, follow these steps:
1. Turn the spanner wrench tool counterclockwise to unscrew the cannula barb.
2. Remove the cannula barb.
3. Check that there is no debris left inside. Insert the new integrated cannula barb and output filter.
4. Turn the spanner wrench tool clockwise until the cannula barb is securely attached. Do not overtighten.
1
10.5 COLUMN CHANGE (RP-502)
The device is programmed to alert you when the columns should be replaced (see 'Alarms' section).
Although you will need to purchase columns from the manufacturer or your service provider, the columns
are designed to be easily changed by the patient by following these steps:
1. Turn off the device by pressing and holding the power button.
2. If using, remove the device from the carry bag or backpack.
3. Remove the battery from the device.
4. Place the device on its side so that the underside is visible.
5. The columns are on one side of the device.
1
©2023 Inogen, Inc. All rights reserved.
2
2
Page 29 of 40
3
3
4
5
96-13115-00-01 B