Error
Number
Error Cause
18
Multiple channels are not allowed to be selected at the same time for
the Digital/Sonic inputs. This error usually means that the sonic port
and a corresponding digital port have been selected.
22
Command 2 contains invalid data.
30
The filter type must be between 0 and 6 for NON-REALTIME data
collection mode and 0, 7, 8, or 9 for real time data collection mode.
This error results from a filter selection outside of this range.
31
Command 3 was sent prior to performing any channel setups.
32
Sample time must be greater than 0 and less than 16000 seconds. The
value is normally rounded to the nearest 100 µsec, but can be rounded
to the nearest 50 µsec in FASTMODE. If the selected channels cannot
support the rate selected, a slower sample rate will be used.
33
The number of samples must be -1 for real time sampling and
between 1 and 12,000 for NON-REALTIME sampling. 0 is not allowed
except for a special case of real time sampling with manual entry.
34
Trigger type must be an integer between 0 and 6. Any other value will
produce this error.
35
The trigger channel must be a valid channel number (e.g., 1-3 or 11)
and must be have been enabled using the channel select command.
36
The trigger threshold must be a value between the maximum and
minimum legal values for the sensor selected. For example, for the
+/-10V probe, legal values are from –10V to +10V.
37
The prestore value must be an integer between 0 and 100%. Any
other values will produce this error message.
38
The external clock parameter is limited to values of 0 or 1. Any other
value will produce this error.
39
The record time parameter is limited to values between 0 and 2. Any
other values will produce this error message.
40
This error will occur when too few parameters are sent in the list. For
example, when setting up an equation with 5 constants, if only 4 are sent,
this error will result.
42
The equation channel number must be a 0 to reset equation, or a 1-3
for the analog channels or 11 for the sonic channel. Equation numbers
outside of this range will produce this error.
43
The equation number must be in the range of .1 to 12 for analog
channels and either 0 or 13 for the sonic channel. Equation numbers
outside of this range will produce this error.
44
The order of the equation must be appropriate for the equation type
selected. For example, an equation order of 5 is not valid for the
mixed polynomial equation.