Syntax: {1, channel,operation,post-processing,(delta)
Use this syntax to set up analog channels.
Channel
1-3 = Analog
0 = Turn channel off
1 = Auto-ID this sensor
(default is 0-5V sensor)
2 = TI Voltage „10V
3 = Current „10A
4 = Resistance
4
5
= Measure period on
„10V input line (CH 1
only)
4
6
= Frequency on „10V
input line (CH 1 only)
5
7
= Count transitions
on „10V input line
(CH 1 only)
10 = TI Temperature or
Stainless Steel
Temperature
(Centigrade)
11 = TI Temperature or
Stainless Steel
Temperature
(Fahrenheit)
12 = TI Light
14 = Low voltage (0-5V)
1
This parameter is ignored.
2
RT: REALTIME mode sampling.
3
NON-RT: NON-REALTIME mode sampling.
4
When using Command 1 operation 5 or 6, Trigger Type in Command 3 must be 2, 3, 4
or 5.
5
When using Command 1 operation 7, Trigger Type in Command 3 must be 0.
24
Operation/
Sensor Type
Post-Processing
0 = None (RT
2
3
and NON-RT
)
1 = d/dt (NON-RT)
2
2
2 = d/dt and d
/dt
(NON-RT)
CBL 2 Technical Reference
1
,equ }
Conversion Equation
0 = Off
1 = On (must also
send Command 4)