Auto-ID Sensors

The CBL 2 contains provisions for the auto-ID sensor resistor values listed below. If
needed, a conversion equation is loaded automatically for some of the auto-ID values.
IDENT
Valu e
1
Sensor Type
2.2K
Thermocouple °C
33K
TI Voltage sensor
6.8K
Current sensor
3.3K
Resistance sensor
22K
Extra long temperature sensor for °C
68K
CO
gas sensor (PPM)
2
100K
Oxygen gas sensor (PCT)
150K
C V voltage sensor (V)
220K
C V current sensor (A)
10K
Stainless steel or TI temperature sensor
15K
Stainless steel or TI temperature sensor for °F
4.7K
TI Light sensor
1K
Ex heart rate sensor (BPM)
47K
Voltage sensor
1.5K
EKG
1
IDENT values are resistance values in ohms (tolerance ±5%).
2
Operation 3 is a mathematical conversion of voltage to a current reading (1V=1A).
There is no circuitry inside the CBL 2 unit to convert current to voltage; this must be
done in the external probe.
3
Default units for the Stainless Steel and TI Temperature sensors is °C.
IDENT
Valu e
1
Sensor Type
15K
Motion detector, meters
22K
Motion detector, meters
10K
Motion detector, feet
33K
Photogate sensor
1
IDENT values are resistance values in ohms (tolerance ±5%).
14
Channels 1, 2, and 3
2
Channel 11 (SONIC)
Range
M200C to 1400¡C
M10 to +10 Volts
M10 to +10 Amps
1K to 100KJ
L50°C to 150°C
0 to 5000 ppm
0 to 27%
M6 to +6 Volts
M0.6 to +0.6 Amps
3
for °C
M25¡C to 125¡C
M13¡F to 257¡F
0 to 1
N/A
0 to 5 Volts
N/A
Range
½ meter to 6 meters
½ meter to 6 meters
1½ feet to 18 feet
N/A
CBL 2 Technical Reference

Advertisement

Table of Contents
loading