Appendix
Caution:
DO NOT use portable radio frequency (RF)
communications equipment (including peripherals such as
antenna cables and external antennas) closer than
12 inches (30 cm) to the Omnipod GO Pod, as degradation of
the performance of this equipment could result.
One exception is that if you use a wearable device, you should
check that device's user guide. If it says the device uses
Bluetooth Wireless Technology, then follow that user guide's
instructions about how close you can place it to other devices.
This equipment is intended for use in the electromagnetic environment specified below.
The user of this equipment should assure that it is used in such an environment.
Compliance
Emissions
according to
RF Emissions
Group 1, Class B
CISPR 11
The Pod is intended for use in the electromagnetic environment specified below. You
should observe these requirements in the use of the Pod.
IEC 60601-
Immunity
1-2
against
test level
ElectroStatic
contact
Discharge, ESD
discharge:
(IEC 61000-4-2)
± 8 kV
air discharge:
± 15 kV
Power
30 A/m
frequency
magnetic fields
50/60 Hz (IEC
61000-4-8)
Radiated RF
10 V/m at
(IEC 61000-4-3)
80 MHz–
2.7 GHz
The table below lists the immunity levels at specific test frequencies
for testing the effects of some wireless communication equipment.
76
Electromagnetic Emissions
Electromagnetic environment
The Pod is suitable for use in all
establishments including domestic
establishments.
Electromagnetic Immunity
Compliance
level (of this
device)
± 8 kV
± 15 kV
400 A/m
10 V/m
Electromagnetic environment
If floors are covered with synthetic
material, try to avoid electrostatic
discharges.
Suitable for most environments.
Magnetic field strengths in excess
of 400 A/m would be unlikely
except in close proximity to
industrial magnetic devices.
Suitable for most environments.
Keep portable RF communications
equipment at least 12 inches (30 cm)
away from the Omnipod GO Pod.