2 Applying or Removing a Pod
2. Push the plunger up to expel any air bubbles out of the fill syringe
and into the insulin bottle.
3. Pull down on the plunger again, if necessary, to continue to fill the
syringe to the "Fill Here" line. Once filled, remove the syringe from
the bottle.

2.9. Fill the Pod

Be sure you've followed all instructions earlier in this chapter before
you proceed to fill the Pod.
Caution:
the fill port and not into any other location
on the Pod. Do not insert the fill syringe
more than once into the fill port. Use only
the fill syringe and needle that came with
your Pod. The fill syringe is intended for
single use only and should be used only with
the Omnipod GO Pod it is packaged with.
Failure to follow these instructions may
result in damage to your Pod.
Caution:
insulin fill port. Insert the fill syringe only straight down into
the insulin fill port. Inserting the syringe at an angle may
damage the syringe or the Pod.
Caution:
feel significant resistance while slowly pressing the plunger
down on the fill syringe. Do not try to force the insulin into
the Pod. Significant resistance may indicate that the Pod has
a mechanical defect. Using this Pod could result in under-
delivery of insulin that can lead to high glucose.
Tip: Read sections 2.9 through 2.14 a few times before you
put on your first Pod. This way you are comfortable not only
with the steps to first fill the Pod, you also understand the
steps to apply the Pod before the 3 minutes have passed after
seeing the amber light.
30
ALWAYS insert the fill syringe into
DO NOT insert the fill syringe at an angle into the
NEVER use a Pod if, while you are filling the Pod, you

Advertisement