Protection from Over-infusion or Under-infusion
The Pod software monitors the infusion rate. If an error that would
result in over- or under-infusion is detected and cannot be corrected,
insulin delivery stops, an alarm sounds and the Pod light shows as
solid red. See "3.4. When the Pod has Stopped: Alarms" on page 48 for
more information about alarms.
Blockage (occlusion) detection
A blockage (occlusion) is an interruption in insulin delivery from the
Pod. A blockage may result from blocked tubing, Pod malfunction, or
from using old or inactive insulin. If the Omnipod GO Pod detects a
blockage, it sounds a hazard alarm and its Pod light turns red to tell
you to change your Pod.
A blockage (occlusion) hazard alarm sounds when an average of
3 units to 5 units of missed insulin occurs. The following table depicts
blockage detection for three different situations when using U-100
insulin. For example, if the Pod's cannula becomes blocked while
using a 25 U/day Pod, 5.5 hours may pass before the Pod sounds a
hazard alarm.
Basal
10 U/day
25 U/day
40 U/day
If a blockage spontaneously clears up, a volume of insulin could
be released. That volume would not exceed the volume of the
pre-programmed insulin intended for delivery.
Time between blockage (occlusion)
and Pod alarm
Typical time
8.5 hr
3.5 hr
2.5 hr
Appendix
Maximum time
12.5 hr
5.5 hr
3.5 hr
69
Need help?
Do you have a question about the Omnipod GO and is the answer not in the manual?
Questions and answers