The graph below each table shows the minimum and maximum
observed errors within the flow accuracy data for any given time
period up to 25 hours. In other words, for a given duration of wear,
the graph shows you the worst potential error in insulin delivery you
could expect.
For the following table and graph, the measurements were made
using a Pod with a preset insulin rate of 10 U/day at a low operating
temperature. The overall mean percentage flow error was 0.90%.
Low Basal Rate Delivery Performance (10 U/day)
Row
Basal
#
Duration
1
Number
of units
requested
2
Amount
Delivered
(median)
3
Amount
Delivered
[min, max]
4
Error %
[min, max]
.
1 hour
0.42 U
0.42 U
[0.14, 0.64]
[-67%, 55%]
EpMax
EpMin
Observation Interval (hr)
6 hours
2.50 U
2.53 U
[1.72, 3.09]
[-31%, 24%]
[-23%, 3.6%]
Appendix
12 hours
5.00 U
5.05 U
[3.83, 5.18]
71