Current Calibration Initialization Commands - Keithley SourceMeter 3A 2420 Service Manual

Table of Contents

Advertisement

2-20
Calibration
4.
5.
Table 2-10

Current calibration initialization commands

Command
:SOUR:FUNC CURR
:SENS:VOLT:PROT 20
:SENS:VOLT:RANG 20
:OUTP:STAT ON
Program the source to output the negative full-range value using the
:SOUR:CURR command. For example:
:SOUR:CURR -1E-3
Note and record the multimeter reading.
Use the multimeter reading as the parameter for the :CAL:PROT:SOUR and
:CAL:PROT:SENS commands. For example, a typical value for the 1mA range
would be:
:CAL:PROT:SOUR -1.025E-3
:CAL:PROT:SENS -1.025E-3
Program the current source for 0A output using the :SOUR:CURR 0.0 command.
Note the multimeter reading.
Send the source and sense calibration commands using the multimeter reading for
the parameter. For example:
:CAL:PROT:SOUR 1E-6
:CAL:PROT:SENS 1E-6
Set the source to the positive full-range value using the :SOUR:CURR command.
For example, for the 1mA range:
:SOUR:CURR 1E-3
Note and record the multimeter reading.
Send the source and sense commands using the multimeter reading as the param-
eter. For example:
:CAL:PROT:SOUR 1.03E-3
:CAL:PROT:SENS 1.03E-3
Send the :SOUR:CURR 0.0 command to set the source current to 0A.
Note and record the multimeter reading.
Send the :CAL:PROT:SOUR command using the multimeter reading as the com-
mand parameter. For example:
:CAL:PROT:SOUR 1E-6
Connect the 1Ω resistor and DMM to the Model 2420 INPUT/OUPUT jacks, as shown
in Figure 2-3. Select the DMM DC volts function.
Repeat step 3 for the 3A range using the calculated current as follows: I = V/R, where V
is the DMM voltage reading, and R is the characterized value of the 1Ω resistor.
Descripton
Select source current mode.
Voltage limit when current source is active.
Make sure 60V range is not active.
Turn source on.

Advertisement

Table of Contents
loading

Table of Contents