Front Panel Voltage Calibration - Keithley SourceMeter 3A 2420 Service Manual

Table of Contents

Advertisement

2-10
Calibration
Table 2-3

Front panel voltage calibration

Source
range*
0.2V
2V
20V
60V
*Use EDIT and RANGE keys to select source range.
**Multimeter reading used in corresponding calibration step. See procedure.
Step 3: Current calibration
Perform the steps below for each current range, using Table 2-4 as a guide.
1.
Figure 2-2
10 µ A to 1A range current
calibration test connections
Source voltage
+200.00mV
+000.00mV
-200.00mV
-000.00mV
+2.0000V
+0.0000V
-2.0000V
-0.0000V
+20.000V
+00.000V
-20.000V
-00.000V
+60.000V
+00.000V
-60.000V
-00.000V
Connect the Model 2420 to the digital multimeter, as shown in Figure 2-2. Select the
multimeter DC current measurement function.
EDIT
V
I
DISPLAY
0
1
TOGGLE
LOCAL
REL
POWER
6
7
DIGITS SPEED
Digital Multimeter
Multimeter voltage
reading**
___________ mV
___________ mV
___________ mV
___________ mV
___________ V
___________ V
___________ V
___________ V
___________ V
___________ V
___________ V
___________ V
___________ V
___________ V
___________ V
___________ V
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
5V
75V
PEAK
PEAK
PEAK
2420 3A SourceMeter
LO
MEAS
SOURCE
250V
PEAK
FCTN
V
I
RANGE
2
3
4
5
EDIT
AUTO
TERMINALS
FILTER
LIMIT
TRIG
SWEEP
RANGE
ON/OFF
FRONT/
8
9
+/-
REAR
STORE RECALL
CONFIG MENU
EXIT
ENTER
OUTPUT
Model 2420
Input LO
Amps