C-2
Calibration Programs
Introduction
This appendix includes a calibration program written in BASIC to help you in calibrate the
Model 2420, as well as an example program that demonstrates how to request calibration con-
stants. Refer to Section 2 for more details on calibration procedures, equipment, and connec-
tions. Appendix B covers calibration commands in detail.
Computer hardware requirements
The following computer hardware is required to run the calibration programs:
Software requirements
In order to use the calibration programs, you will need the following computer software:
Calibration equipment
The following calibration equipment is required:
See Section 2 for detailed equipment specifications.
General program instructions
1.
2.
3.
IBM PC compatible computer.
Keithley KPC-488.2, KPS-488.2, or KPC-488.2AT, or CEC PC-488 IEEE-488 interface
for the computer.
Two shielded IEEE-488 connecting cables (Keithley Model 7007)
Microsoft QBasic (supplied with MS-DOS 5.0 or later).
MS-DOS version 5.0 or later.
HP-style Universal Language Driver, CECHP.EXE (supplied with Keithley and CEC
interface cards listed above).
Hewlett-Packard HP3458A Digital Multimeter
1Ω, ±0.1%, 100W resistor
With the power off, connect the Model 2420 and the digital multimeter to the IEEE-488
interface of the computer. Be sure to use shielded IEEE-488 cables for bus connections.
Turn on the computer, the Model 2420, and the digital multimeter. Allow the Model 2420
and the multimeter to warm up for at least one hour before performing calibration.
Make sure the Model 2420 is set for a primary address of 24. (Use the front panel
MENU/GPIB selection to check or change the address.)