Keithley SourceMeter 3A 2420 Service Manual page 119

Table of Contents

Advertisement

Example
Table B-3
Recommended:CALibration:PROTected:SOURce parameter ranges
Source
First parameter
range
(negative full scale)
0.2V
-0.18 to -0.22
2V
-1.8 to -2.2
20V
-18 to -22
60V
-50 to -65
10µA
-9E-6 to -11E-6
100µA
-90E-6 to -110E-6
1mA
-0.9E-3 to -1.1E-3
10mA
-9E-3 to -11E-3
100mA
-90E-3 to -110E-3
1A
-0.9 to -1.1
3A
-2.7 to -3.15
Note: Actual allowable parameter ranges are ±50% of full range for zero parameters, 75% of full scale to 150% of full
scale for full-range parameters, but ranges listed above should be observed for optimum calibration.
Note that parameters for a given source function and range may be sent in
any order; however, once one :CAL:PROT:SOUR command executes, the
other three must also be executed, or an error number -200, "Execution error"
will occur when the :CAL:PROT:SAVE command is processed at the end of
the calibration procedure.
Because the source is calibrated for both positive and negative values, two
zero calibration points are required. The Model 2420 automatically uses the
appropriate zero parameter based on whether the source polarity is positive or
negative.
:CAL:PROT:SOUR 2
Second parameter
(negative zero)
-0.002 to +0.002
-0.02 to +0.02
-0.2 to +0.2
-0.5 to +0.5
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
-3E-2 to +3E-2
Command Reference
Calibrate 2V source range.
Third parameter
Fourth parameter
(positive full scale)
(positive zero)
+0.18 to +0.22
-0.002 to +0.002
+1.8 to +2.2
-0.02 to +0.02
+18 to +22
-0.2 to +0.2
+50 to +65
-0.5 to +0.5
+9E-6 to +11E-6
-1E-7 to +1E-7
+90E-6 to +110E-6
-1E-6 to +1E-6
+0.9E-3 to +1.1E-3
-1E-5 to +1E-5
+9E-3 to +11E-3
-1E-4 to +1E-4
+90E-3 to +110E-3
-1E-3 to +1E-3
+0.9 to +1.1
-1E-2 to +1E-2
+2.7 to +3.15
-3E-2 to +3E-2
B-7