Purge the sensor case

Prerequisites
Make sure the following are available before beginning the purge procedure:
Teflon
Argon or nitrogen gas sufficient to purge the sensor case
Whenever a purge plug is removed from the sensor case, you must purge the case again.
Procedure
1. Shut down the process, or set control devices for manual operation. Before
2. Remove both purge plugs from the sensor case. If purge lines are being used, open
3. Prepare the purge plugs for reinstallation by wrapping them with 2–3 turns of
4. Connect the supply of nitrogen or argon gas to the inlet purge connection or open
Installation Manual
tape
performing the case purging procedure, shut down the process or set the control
devices for manual operation. Performing the purge procedure while the flowmeter
is operating could affect measurement accuracy, resulting in inaccurate flow
signals.
the valve in the purge lines.
WARNING
Stay clear of the rupture disk pressure relief area. High-pressure fluid escaping
from the sensor can cause severe injury or death. The sensor must be oriented
so that personnel and equipment will not be exposed to any discharge along
the pressure relief path.
Take all necessary precautions when removing purge plugs. Removing a purge
plug compromises the secondary containment of the sensor and could expose
the user to process fluid.
Improper pressurization of the sensor case could result in personal injury.
NOTICE
Be sure to use thread protectors when removing the purge fitting so as not to
damage the disk membrane surrounding the rupture disk.
Teflon tape.
inlet purge line. Leave the outlet connection open.
Exercise caution to avoid introducing dirt, moisture, rust, or other contaminants
into the sensor case.
If the purge gas is heavier than air (such as argon), locate the inlet lower than the
outlet, so that the purge gas will displace air from bottom to top.
If the purge gas is lighter than air (such as nitrogen), locate the inlet higher than
the outlet, so that the purge gas will displace air from top to bottom.
Purge the sensor case
August 2019
21