Potentiometers and Reference Switches
The system may now be moved outside the "Litho Position". (X = -130 / + 70; Y = ± 320)
Safety limit switches S 76/77 as well as S 78 must not be activated.
(Refer to circuit diagram J1042-7).
Remove the Uro insert and move to the →0← basic position.
The system will only allow travel in the "Litho Position".
Move the W drive out of the park position.
Switch S79 must open in the MC position - 430 ± 2 mm and switches S 76; S77; S78
must be activated. If necessary, check "Switch ON" with the DVM on board D32 X41
PIN 3 and at ground GND.
Siemens AG