Mains connection and power supply
Checks before powering up the generator
On p.c. board D702:
1
Set switch S1(UZW) to "OFF" (lower position).
2
Set switch S2 (SS) to "OFF" (lower position).
3
Set switch S3 (TEST) to "ON" (upper position- output service PC).
S1
UZW
SS
TEST
V6
V7
ZBL
V1
AR
V2
KVE
V3
BRAKE
V4
STATUS
V5
VH
X967
NSE1
X980
NSE2
X981
DS
X982
SS
X983
BS
X984
GND
X985
X990
Figure 1 P.C. board D702
Measuring the line resistance
See Figure 2
1
Remove fuses F10 (1) and F20 (2) or, when single-phase supply, fuse F10 (1).
2
Connect line resistance meter to fuse holders F10 and F20.
3
Mains supply "ON".
4
Carry out measurement.
5
Mains supply "OFF".
Siemens-Elema AB
Solna, Sweden
J39
S2
S3
X962
X964
X960
X963
R13
R11
R43
SPS-UD
RX B7-120.033.01.10.02
J50
X965
IHREG
X968
X961
X969
MAREG
05.98
X966
X702
MAMMOMAT 300
65 02 186
7-1