Siare SIARETRON 1000 Service Manual page 39

Table of Contents

Advertisement

SIARETRON 1000 – 1100 - 3000
Service Manual
19) Verify the supply gas alarm
Set:
-
IPPV operative mode.
%insp. = 50.
-
Connect the patient circuit to simulator.
Close AIR supply on the couple.
Verify that an alarm sounds, that the Gas Supply led lights-on and that the A04 script is flashing on Minute
Volume display.
Restore the AIR supply.
Close the O
supply on the couple.
2
Verify that an alarm sounds, that the Gas Supply led lights-on and that the A03 script is flashing on Minute
Volume display.
Restore the O
. supply.
2
20) PEEP control
Set:
-
IPPV operative mode.
Connect the patient circuit to simulator.
Rotate the PEEP regulator, until 10 cmH
Verify that the pressure on simulator manometer is of 10 cmH
Rotate the PEEP regulator to the max. value and verify that on the Max/Mean Airway Pressure display is
shown the 20 cmH
2
21) Verify the ventilator functionality
Set:
- %INSP = 33%.
Select the IPPV+ASS operative mode by pressing the corresponding button on the front panel.
Connect the patient circuit to the Low flow range inlet of RESPICAL which is setted in "VOLUME" modality and
option "La".
Set the following MINUTE VOLUME values and for each of them the corresponding RATE value as indicated in
the table; verify that the TIDAL VOLUME values displayed on RESPICAL are within the corresponding
acceptance limits (see table).
NOTE: Setting the Minute Volume at 15 and 25 it is necessary to pass to the HIGH FLOW inlet on the
RESPICAL always in "VOLUME" modality and option "La".
SIARETRON_SE3_en
O is shown on Max/Mean Airway Pressure display.
2
O value and that the pressure on simulator manometer is of 20 cmH
O ± 1 cmH
2
39 / (72 )
O.
2
O ± 2 cmH
2
( rev.3) 30.06.2004
O.
2