Siare SIARETRON 1000 Service Manual page 37

Table of Contents

Advertisement

SIARETRON 1000 – 1100 - 3000
Service Manual
13) Verify the power failure (power failure alarm)
Set the IPPV+ASS. operative mode.
Disconnect the electric main power supply.
Verify that:
-
the ON LINE led switches-off and the ON-BAT led switches-on flashing
-
an acoustic alarm occurs
-
the alarm is resettable by pressing the alarm reset button
-
the ventilator keep on working
-
the Rate, Minute volume display and airways pressure led bar are always on
-
every 10 seconds are displayed also all other setted parameters.
!
14) Verify the bronchomanometer
Connect the patient circuit by a T connector to the patient simulator and to the P+ inlet of RT200 setted with
program 12 and with PEAK option.
Set the following parameters:
-
operative modality = IPPV
-
Flow = 10 L/min ( to check the PEEP at 20 ) – 5 L/min (to check the PEEP at 10 );
-
Rate = 20;
Act on PEEP knob so to set on Max/Mean airway pressure display the following pressures and verify that
the displayed values on RT200 are within the corresponding acceptance limits.
Setted ventilator PEEP
20
10
!
15) Verify leakage absence
15a) Tightness test from valves group to patient (Low pressure circuit leakage)
Connect the patient circuit to patient simulator.
Set the following parameters:
- Operative modality = SPONT SIMV;
Act on PEEP knob so to set on airway pressure bar led the pressure of 20 cm H 2 O.
Verify that on the airway pressure bar led and on patient simulator manometer is displayed the pressure of
O ± 2 cmH
20 cmH
2
15b) Tightens Test of circuit (Leakage on main circuit)
Switch-off the ventilator.
Close upstream the manometer on the couple gas supply.
Verify that the pressure on manometer does not decrease under 3 bar in 30 sec.
SIARETRON_SE3_en
-%INSP = 33.
Lower acceptance limit
O for all the setted time (1 minute).
2
(read on RT200)
18
9
- SIMV RATE = 1;
37 / (72 )
Upper acceptance limit
(read on RT200)
22
11
( rev.3) 30.06.2004

Advertisement

Table of Contents
loading

This manual is also suitable for:

Siaretron 1100Siaretron 3000

Table of Contents