Final Testing - Siare SIARETRON 1000 Service Manual

Table of Contents

Advertisement

SIARETRON 1000 – 1100 - 3000
Service Manual
4.6

FINAL TESTING

Testing equipment:
couple 220Vac Power supply cable
ATS.033 patient simulator
RT 200
paediatric patient circuit of 120 cm
1)
Effect the following connections:
-
connect the electric power cable to ventilator and to main power supply;
-
connect to the couple compressed air and oxygen at 3,5 bar, connect the couple exit cables
respectively to the AIR and O
-
connect the paediatric patient circuit of 120 cm to ventilator and to ATS.033 patient simulator.
2)
Verify the marks presence and check that the following labels are present:
2a) Identification label with data relevant to power supply predisposed for the "equipment";
2b) label WARNING DO NOT USE IN PRESENCE OF INFLAMMABLE ANAESTHETIC GAS, in Italian and
in English.
!
3)
Verify fan wheel alarm.
Effect the following checks:
3a) with moving fan wheel the alarm must not occur;
3b) with disconnected fan wheel the alarm must occur;
3c) with mechanically blocked blades the alarm must occur.
4)
Autotest 1 and Autotest 2
Switch-on the ventilator. Immediately after the start-up the machine AUTO-TEST 1 begins.
Verify that the autotest is positive, namely that no error script appears (Exx or Axx) in the MINUTE
VOLUME display on the front panel.
Check that at the end of calibration the machine working hours number is displayed on the Minute Volume
display.
At the end of autotest the machine must go automatically in STANDBY (all operative modes leds are off).
Switch-off and on the ventilator keeping pressed the CPAP button (operative mode), so that immediately
the machine AUTO-TEST 2 starts.
Verify that the autotest is positive, namely that no error script appears (Exx or Axx) on MINUTE VOLUME
display on the front panel.
Check that at the end of calibration the machine working hours are displayed on the Minute Volume
display.
At the end of autotest the machine must go automatically in STANDBY (all the operative modes leds are
off).
SIARETRON_SE3_en
lung ventilator inlets;
2
33 / (72 )
( rev.3) 30.06.2004