SIARETRON 1000 – 1100 - 3000
Service Manual
Testing Performance Procedure
Acting on the PANEL board, while the alternated scripts: I-O | ofG (on DS13/14/15) appear, perform the
following checks (the referrals under brakes are relevant to the positions on the PANEL board):
By the AUT. TIME CYCLED (SW7) button the ON/OFF electrovalve must light ON and OFF
following the status of the LED (LD1) next to the (LED ON, LED OFF) button.
By the AUT. PRESS CYCLED (SW2) button the BUZZER must light ON and OFF following the
status of LED (LD2) next to the (LED ON, LED OFF) button.
With the main power supply the green ON LINE LED (LD12) must be light on.
With the external battery the yellow ON BATT LED (LD11) must be light on, simulate the
external battery with power supply regulated at 12 Vdc connected as follows:
Verify the connection of LOW BATT (LD10) alarm LED lowering the external power supply
whom simulates the battery. At an electric power between 10.5 and 11.2Vdc the alarm must
intervene.
By pressing the ON/OFF switch (SW1) the yellow SPONT. BREATH DETECTOR LED (LD5)
must light on.
Short-circuiting the CN6 connector feet of the LV200I/O board (Air inlet pressure switch) the red
LOW AIRWAY PRESS (APNEA).LED (LD7) must light off.
Short-circuiting the CN7 connector feet of LV200I/O board (O2 inlet pressure switch) the red
GAS SUPPLY.LED (LD9) must light off.