Press at the same time the two LOW AIRWAY PRESS.
light-on, hence regulate their light intensity by mean of the trimmers present on the boards (as indicated here
below):
Press any switch to return to scrolling script.
2)
Functional checks:
2.a)
Pressing the ALARM RESET key it is possible to effect the LED BAR check.
The leds are lighted on one by one in sequence: check that all leds light-on and that only one is on at a time.
2.b)
Pressing again the ALARM RESET key it is possible to verify the SINGLE LEDS
The leds are lighted-on one by one in sequence: check that al leds light-on and that only one is
on at a time.
2.c)
Pressing again the ALARM RESET keyit is possible to verify the KEYBOARD.
On MINUTE VOLUME display message " tAStiErA " scrolls, while on MAX/MEAN AIRWAY
PRESS display the SCAN CODE OF PRESSED SWITCH APPEARS (0 for any switch). Verify
the correspondence of the switches with the following table:
SCAN CODE
SIARETRON_SE3_en
Light Iny Regulation
General
1
CPAP
2
SPONT.SIMV
3
AUT.PRESS.CYCLED
4
AUT.TIME CYCLED
5
SIMV RATE
6
SIMV RATE
7
02 CONC
8
02 CONC
9
LOW AIRWAY PRESS.
and
switches; verify that all panel displays and leds
R7
Led bar+ Single Leds+ Display at
MAX AIR MIN VOL
R6
R21
KEY
26 / (72 )
7 segments
R12
SWITCH
SW4
SW3
SW2
SW7
SW6
SW5
SW10
SW9
SW8