At each data confirmation a beep sounds and the new flow value to be searched is displayed; after the last
confirmed value a flag in e2prom is resetted (= valid tables) and the automatic procedure starts.
During such procedure the same previously described data are displayed; only if the calibration is correctly
carried out, previously resetted flag is restored, otherwise the machine cannot work and at any start-up an E16
error message is generated.
NOTE: the performance of the whole calibration procedure of proportional electrovalves takes approx. 3 hour
and it is not unterruptable; without its correct completion the ventilator signals at any start-up the Er16 error
message and it does not allow the normal operation. Switch-off the ventilator
13)
The duration Test: connect the ventilator to pneumatic power supply and to a patient circuit with a small
simulation bag. Switch-on the ventilator and after the autotest phase, take it to the automatic operation
with the following parameters:
-
RATE = 45 bpm
-
MINUTE VOLUME = 1,7 l/min
-
O2 CONC: = 21%
-
I:E RATIO = 33%
Put the max. and min. pressure limits at 5 - 7 cmH
ventilator in continuous operation for at least 24 h. At finishing of this test verify that any alarm condition
is activated. After that switch-off the ventilator.
NOTE: Position the bag/simulator so that it does not move during all test phase. Every movement of
bag/simulator will cause a peak pressure variation, leading to the alarm condition.
14)
Zeroing of working hours : press the on/off switch and keeping pressed the LOW AIRWAY PRESS.
and HI AIRWAY PRESS
the CPAP button within 5 seconds . The "partial" worked hours and minutes are scrolled on the (namely
counted from the last zeroing) MINUTE VOLUME display in format: h. xxx Min. xx . In this procedure
the following keys are active:
HI AIRWAY PRESS
HI AIRWAY PRESS
AUT. TIME CYCLED + RATE
(simultaneously pushed)
CPAP + TRIGGER SENSITIVITY
(simultaneously pushed)
ALARM RESET
Verify that the worked hours are more than 24 and zeros the "partially" worked hours ( AUT. TIME
CYCLED + RATE
SIARETRON_SE3_en
keys until a short sound, so release the above mentioned keys and press
Displays the "historical" sum of the worked hours in the
following format: tot. h. xxx.xxx Min. xx . Such a sum considers
also the registered "partially" worked hours.
Displays the "partially" worked hours in the previously described
format.
Zeroing of "partially" worked hours after having added them to
the "historical" hours.
Zeroing of "historical" sum of worked hours.
Terminates the service procedure and makes the autotest start.
buttons) and totals (CPAP. TRIGGER SENSITIVITY
32 / (72 )
O over and under the pressure peak. Leave
2
buttons).
( rev.3) 30.06.2004