Siare SIARETRON 1000 Service Manual page 27

Table of Contents

Advertisement

SIARETRON 1000 – 1100 - 3000
Service Manual
SCAN CODE
NOTE: the on/off switch has not a scan code because its contact is directly taken on the POWER integrated
circuit.
In order to pass to the next phase it is necessary to press the two ALARM RESET and CPAP switches at the
same time.
2.d)
In this last phase the correct start-up of all 7 segments displays leds must be controlled.
In all displays the segments light on in sequence as follows: a-b-c-d-e-f-g-dp (the latter one if
connected).
The leds are lighted-on one by one in sequence: verify that all leds light-on and that only one is
on at a time
In all groups the more significative number is to be considered first; the switch-on order of displays is the
following:RATE-%INSP-TRIGGER SENSITIVITY-02 CONC.-SIMV RATE-MAX/MEAN AIRWAY PRESSURE-
MINUTE VOLUME. Only the last two groups (namely MAX/MEAN AIRWAY PRESSURE and MINUTE
VOLUME) have the connected dp.
After scanning the scrolling script returns: "PAnnEL Lu200 rEL. 4.00".
SIARETRON_SE3_en
10
- LOW AIRWAY PRESS.
11
- HI AIRWAY PRESS.
12
HI AIRWAY PRESS.
13
RATE
14
RATE
15
%INSP.
16
%INSP.
17
TRIGGER SENSITIVITY
18
TRIGGER SENSITIVITY
19
ALARM RESET
f
e
d
TASTO
a
b
g
c
27 / (72 )
SWITCH
SW13
SW12
SW11
SW16
SW15
SW14
SW17
SW18
SW19
SW20
( rev.3) 30.06.2004