Overview; Introduction; Specifications - Chroma A195001 User Manual

Pd calibrator
Table of Contents

Advertisement

1.

Overview

1.1 Introduction

The A195001 PD Calibrator is designed with reference to the IEC60270-1 calibration
standards, which can be used in daily check to verify the measurement errors or calibrate the
PD tester.

1.2 Specifications

Item
Charge Range
Accuracy
Rise Time
Pulse Repetition
Power Supply
Current Consumption
Dimension
Net Weight
Operable Range
Storage Range
Specification
1.0, 2.0, 5.0, 10.0, 20.0, 50.0, 100.0 pC
100pC
Injection Capacitance: 1pF, typical
20, 50, 100, 200, 500, 1000, 2000 pC
2000pC
Injection Capacitance: 20pF, typical
Polarity
Positive, Negative
± (3% of setting + 0.5pC )
< 50ns
100Hz
9V battery
50mA max.
65(W) x 150(H) x 36.5(D) mm
Approx. 500g
− Temperature and Humidity Range: 0°C to 45°C,
15% to 95% and no condensation.
− Pollution degree rating: Pollution Degree 2. Rated
for indoor, dry location use only.
− Working altitude: Up to 2000 meters.
-10 to 50°C ,  95% RH
Overview
1-1