7 Adjustment

As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization.
Press
and
mode. [ECF 1] is displayed.
(The adjustment process can be exited at any time with
then to weighing mode.)
Confirm display of [ECF 1] with
displayed.
Acknowledge with
be displayed briefly. Then the display to input the
recommended adjustment weight is displayed
(see chap. 1 "Technical data")
The left digit flashes.
Enter the value of the adjustment weight as follows:
Switch to the next digit to the right with
Increase number with
Confirm the value entered with
18
at the same time in weighing
. [CAL Z] is
. [--------] followed by [LOAd
.
.
.
. The balance returns
i
] will
(example)
EOB / EOE / EOS -BA-e-1741