Quick Setup - Endress+Hauser SmarTec S CLD 132 Operating Instructions Manual

Conductivity measuring system
Hide thumbs Also See for SmarTec S CLD 132:
Table of Contents

Advertisement

SmarTec S CLD 132
Endress+Hauser
Alarm contact
A
ˆ}°ÊÎä’Ê Di†œ““i˜–i–Êvbˆ‡ÃbviʆˆÀ†ÕˆÌÊvœÀÊb˜ÊbbÀ“ʆœ˜Ìb†Ì
Á 2œÀ“bÊœ«iÀb̈˜}ÊÃÌbÌi
Normal operating state
• Instrument in operation
• No error message available (Alarm
LED off)
➝ Relay picked up
➝ Contact 42/43 closed
6.3

Quick setup

After switching the transmitter on, configure the major functions required for accurate
measurement. The following section gives you an example for a basic configuration.
Input
1. Press the ENTER key.
2. Enter the code 22 to be able to edit the setup.
Press the ENTER key.
3. Press the MINUS key several times until the
"Service" function group is displayed.
4. Press the ENTER key to edit this function group.
5. Select your language, e.g. "ENG" for English.
Confirm your entry be pressing the ENTER key.
6. Press the PLUS and MINUS keys simultaneously
to quit the "Service" function group.
7. Press the MINUS key several times until the
"Setup 1" function group is displayed.
8. Press the ENTER key to edit "Setup 1".
9. In A1, select the operating mode, e.g.
"cond" = conductivity.
Confirm your selection by pressing the ENTER
key.
41
42
43
É ÁbÀ“ÊÃÌbÌi
Alarm state
• Error message available (Alarm LED red)
or
• Instrument defective or voltage-free (Alarm
LED off)
➝ Relay dropped out
➝ Contact 41/42 closed
Selection or range
(factory setting
bold)
ENG = English
GER = German
FRA = French
ITA = Italian
NEL = Dutch
ESP = Spanish
cond = conductivity
conc =
concentration
Commissioning
B
41
42
43
C07-CLD132xx-04-06-00-xx-005.eps
Display
S
SERVICE
ENG
S1
Language
A
SETUP 1
cond
A1
Oper.Mode
29

Advertisement

Table of Contents
loading