6.3

Quick Setup

In the case of measuring devices without a local display, the individual parameters and functions
must be configured via the configuration program, e.g. FieldCare. If the measuring device is
equipped with a local display, all the important device parameters for standard operation can be
configured quickly and easily by means of the "Commissioning" Quick Setup menu.
• Quick Setup "Commissioning", see below
• Quick Setup "Pulsating Flow" → ä 43
• Quick Setup "Gas measurement"→ ä 46
6.3.1

Quick Setup "Commissioning"

Note!
• The display returns to the function SETUP COMMISSIONING (1002) if you press the Q key
combination during parameter interrogation. The stored parameters remain valid.
• The "COMMISSIONING" Quick Setup must be performed before another Quick Setup is run.
m
The "DELIVERY SETTINGS" option sets every selected unit to the factory setting.
The "ACTUAL SETTING" option accepts the units you previously configured.
n
Only units not yet configured in the m current Setup are offered for selection in each cycle. The unit for mass,
volume and corrected volume is derived from the corresponding flow unit.
o
The "YES" option remains visible until all the units have been configured.
"NO" is the only option displayed when no further units are available.
p
The prompt only appears if a current and/ or pulse/frequency output is available. Only the outputs not yet
configured in the current Setup are offered for selection in each cycle.
q
The "YES" option remains visible until all the outputs have been configured.
"NO" is the only option displayed when no further outputs are available.
r
The "automatic parameterization of the display" option contains the following basic settings/factory settings:
YES
Main line = Mass flow
Additional line = Totalizer 1
Information line = Operating/system conditions
NO
The existing (selected) settings remain.
s
The process for executing other Quick Setups is described in the following chapters.
Commissioning
41