Quick Setup - Endress+Hauser Smartec S CLD134 Operating Instructions Manual

Conductivity measuring system
Hide thumbs Also See for Smartec S CLD134:
Table of Contents

Advertisement

Smartec S CLD134
Endress+Hauser
6.3

Quick setup

After switching the transmitter on, configure the major functions required for accurate
measurement. The following section gives you an example for a basic configuration.
Input
1. Press the ENTER key.
2. Enter the code 22 to be able to edit the setup. Press
the ENTER key.
3. Press the MINUS key several times until the "Service"
function group is displayed.
4. Press the ENTER key to edit this function group.
5. Select your language, e.g. "ENG" for English.
Confirm your entry be pressing the ENTER key.
6. Press the PLUS and MINUS keys simultaneously to
quit the "Service" function group.
7. Press the MINUS key several times until the "Setup 1"
function group is displayed.
8. Press the ENTER key to edit "Setup 1".
9. In A1, select the operating mode, e.g.
"cond" = conductivity.
Confirm your selection by pressing the ENTER key.
10. In A2, press the ENTER key to confirm the factory
setting.
11. In A3, press the ENTER key to confirm the factory
setting.
12. In A4, press the ENTER key to confirm the factory
settting.
13. In A5, enter the cell constant for the connected
sensor. Refer to the sensor's or the compact version's
quality certificate for the exact value.
Selection or range
Display
(factory setting bold)
SERVICE
ENG = English
GER = German
FRA = French
ITA = Italian
Language
NEL = Dutch
ESP = Spanish
SETUP 1
cond = conductivity
conc = concentration
Oper.Mode
%
ppm
mg/l
TDS = Total Dissolved
Conc.Unit
Solids
none
XX.xx
X.xxx
XXX.x
Format
XXXX
auto, μS/cm, mS/cm,
S/cm, μS/m, mS/m,
S/m
Unit
0.10 ... 6.3 ... 99.99
Cellconst
Commissioning
S
ENG
S1
A
cond
A1
ppm
A2
XX.xx
A3
auto
A4
6.300
1/cm
A5
35