Maintenance and Service of your Thermo Refrigerator
All laboratory refrigerators feature automatic defrosting and, in normal conditions, means
that no manual defrosting is required. However, in certain conditions (e.g. when the
refrigerator is very full or when any new items of are stored in the refrigerator) frost may
form on the rear wall of the equipment as the compressor must run for long periods of time.
As automatic defrosting takes place while the compressor is not running the refrigerator may
not defrost completely before the compressor restarts. In these instances some manual
defrosting may be required.
Before manually defrosting the cabinet, make sure the power lead is disconnected from the
mains. Any loose frost found on the evaporator can be removed carefully using a wooden or
plastic scrapper.
WARNING: Do not use any knives or sharp metallic objects to remove frost.
Alternatively a non-scheduled defrost can sometimes be carried out as follows:
Function
To carry out a non‐scheduled defrost
N.B. This does not replace the automatic defrost function on relevant models
• Your refrigerator should be cleaned regularly.
• Before cleaning you MUST switch off and disconnect your refrigerator from the
power supply.
• The whole refrigerator except the door gasket can be cleaned with mild cleaning
detergent.
• The door gasket should be cleaned with water only and wiped dry.
• The handle should be cleaned regularly with soap and water.
• Remove shelves/basket/door furniture and wipe with a soft cloth. DO NOT put then
in a dishwasher.
• Detergents containing abrasives or acids are not suitable for the cleaning.
• When all cleaning has been completed and accessories replaced, reconnect the plug
with dry hands.

Defrosting

Press
followed by
followed by
followed by

Cleaning

Hold button(s)
Example screen
down for
3 seconds
Four times
1 second
1 second