Thermo Scientific Fisher Scientific Isotemp 50FREEFSA Installation And Operational Manual

Thermo Scientific Fisher Scientific Isotemp 50FREEFSA Installation And Operational Manual

Flammable materials storage refrigerator

Advertisement

Quick Links

Installation and Operational Manual
Isotemp™ Flammable Materials
Storage Refrigerator
Model 50FREEFSA

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Fisher Scientific Isotemp 50FREEFSA and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Thermo Scientific Fisher Scientific Isotemp 50FREEFSA

  • Page 1 Installation and Operational Manual Isotemp™ Flammable Materials Storage Refrigerator Model 50FREEFSA...
  • Page 2 IMPORTANT Read this instruction manual. Failure to follow the instructions in this manual can result in damage to the unit, injury to operating personnel, and poor equipment performance. CAUTION All internal adjustments and maintenance must be performed by qualified service personnel. Material in this manual is for informational purposes only.
  • Page 3: Table Of Contents

    Contents Safety Information ............1 Alert Signals ............. 1 Unpacking ..............2 Visible Loss or Damage ..........2 Concealed Loss or damage ........2 Performance Characteristics ........3 Temperature Ranges..........3 Electrical Requirements ..........3 Installation..............4 Selecting a Location..........4 Leveling the Unit............
  • Page 4: Safety Information

    Safety Information Your satisfaction and safety are important to the Fisher Shock Hazard: Use of this equipment involves Scientific channel and a complete understanding of this unit is power supplies which convert line voltage to low necessary to attain these objectives. voltage power.
  • Page 5: Unpacking

    Unpacking Save all packing material if apparatus is received damaged. This merchandise was carefully packed and thoroughly inspected before leaving our factory. Responsibility for its safe delivery was assumed by the carrier upon acceptance of the shipment; therefore, claims for loss or damage sustained in transit must be made upon the carrier by the recipient as follows: Visible Loss or Damage...
  • Page 6: Performance Characteristics

    Performance Characteristics Temperature Range 1°C to 12°C (34°F to 53.6°F) Electrical Requirements 115 Volts ±10%, 60 Hz, 9.1 Amps Isotemp Flammable Materials Storage Refrigerator Performance Characteristics |...
  • Page 7: Installation

    Installation Selecting a Location Choose a location for the refrigerator that will provide at least three inches of clearance between the cabinet and any adjacent vertical surface at the sides and rear. Twelve inches of clearance is required above the grille. Appropriate electrical power must be available.
  • Page 8: Electrical Connection

    • Cable clamp and other equipment on this circuit not exceed the rating of the fuse or circuit breaker in use. WARNING: For personal safety this unit must be properly grounded. The frequency and nominal voltage requirements for the unit are specified on the data plate, which is located on the interiors upper left side.
  • Page 9: Operation

    Operation Initial Startup Material Compatibility Turn power on and verify that the condenser fan is running. The interior cabinet of this unit is constructed of anodized This model has a condenser fan located on the condensing aluminum. Care must be exercised when determining which unit on top of the cabinet.
  • Page 10: Troubleshooting

    Troubleshooting This table is intended to assist in resolving user-correctable Refrigerator problems by relating symptoms to their likely causes.If services beyond the scope of this table is required, contact Customer Service at 1-800-438-4851. Symptom Probable cause of Problem Action Does Not Run Unit Unplugged Plug in Unit Blown fuse or tripped circuit breaker...
  • Page 11: Maintenance

    Maintenance Cabinet Cleaning WARNING: When servicing the unit, disconnect from the electrical power source. CAUTION: Do not use any type of abrasive such as steel wool, or fluids such as gasoline, Naphtha, and thinner.These materials could be harmful to aluminum, plastic materials, door gasket, and painted surfaces.
  • Page 12: Warranty

    Warranty Laboratory instruments and equipment manufactured by e. Representations and warranties made by any person, Fisher Scientific Company L.L.C. – Laboratory Equipment including dealers and representatives of the Company, Division (hereinafter called “the Company”) are warranted only which are inconsistent or in conflict with the foregoing as stated below.
  • Page 13 Important For your future reference and when contacting the factory, please have the following information readily available: Model Number : Serial Number : Date Purchased : The above information can be found on the dataplate attached to the equipment. If available, please provide the date purchased, the source of purchase (manufacturer or specific agent/rep organization), and purchase order number.
  • Page 14 Fisher Scientific Company L.L.C 275 Aiken Road Asheville, NC 28804 United States 104519 Rev.C July 2018 In the United States: For customer service, call 1-800-766-7000 To fax an order, use 1-800-926-1166 To order online: fishersci.com © 2018 Thermo Fisher Scientific Inc. All rights reserved. Trademarks used are owned as indicated at fishersci.com/trademarks.